SOFT K
The Eni-Eye Soft K lens is a soft contact lens especially designed to fit Keratoconus cases. The base curve geometry is made up of a spheric optic zone and an aspheric periphery. The front curve has three zones, a thick optic zone to improve the optic quality, a special lenticular zone for structural stabilization, and an edge zone.
Additional Features
Additional features of the Eni-Eye Soft K Lens include:
- Pressure balancing holes-A special system to equalize the pressure between the front and back of the lens
- Back Toric design of up to – 1.75 DS cyl in any axis.
Application
Correction of Keratoconus
Les Material
The Eni-Eye Soft K (Xylofilcon A) (hydrophilic) Contact Lens for Keratoconus for Daily Wear lens material (Xylofilcon A) is a hydrophilic random copolymer of N-vinyl pyrrolidone, methyl methacrylate and cyclohexyl methacrylate 2-ethoxyethyl methacrylate cross-linked with allyl methacrylate with no color additives or UV absorber. When hydrated, the lens consists of 33% (Xylofilcon A) and 67% water by weight when immersed in normal saline @ 20°C.
| PHYSICAL PROPERTIES |
| Water Content |
67.5% in saline at 20°C |
| Oxygen Permeability |
29 x 10-11 at 35°C, DK (Fatt Scale) |
|
Lens Parameters
The Eni-Eye Soft K is available in the following parameters:
| LENS PARAMETERS (SPHERIC DESIGN) |
| Base Curves |
7.30, 7.60, 7.90 & 8.20mm |
| Diameter |
14.20mm |
| Sphere Power |
+ 4.00 to –12.00 DS |
| Optic Zones |
8.30mm (-3.00 DS) |
| Center Thickness |
0.38mm (-3.00 DS) |
| Edge Thickness |
0.11mm |
|
| LENS PARAMETERS (TORIC DESIGN) |
| Cylinder Power |
-0.50 to –2.00 DS |
| Axis |
Any |
|
Fitting Procedure
- Trial lens set includes 3 lenses 7.30 , 7.60 , and 7.90
- Patient history
- Corneal Topography if possible
- Routine slit lamp eye check
- First choice lens 7.60, wait 30 minutes and check:
- Movement must be 0.7mm to 1.0mm
- The lens must be central
- The patient feels comfortable with the lens
- No Air bubbles
- No edge lift
- Acceptable vision with over refraction
- Over refraction SPH. & CYL. (CYL. Up to –2.00DS)
| Note: An auto refractor gives a false reading over this lens. Use lenses for over refraction. |
|
|
Cleaning
Hydrogen Peroxide or a chemical disinfecting system (MPS) is recommended for cleaning the lens.
|
 |